*** MIDAS logfile (vers. 99NOV) *** We, 21 Mar 2001 16:40:24 Page 1 *** MIDAS logfile (vers. 99NOV) *** We, 21 Mar 2001 16:40:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010321 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.51 stddev = 2.49 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010321 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1753.7 1753.8 809.4 0.08 0.08 2.17 2 8.23 3451.1 3452.2 1553.9 -0.04 -0.01 2.22 3 16.47 6852.6 6849.1 3114.1 -0.13 -0.18 2.20 4 24.70 10251.1 10249.1 4669.4 -0.15 -0.16 2.20 5 37.04 15351.4 15352.4 6990.2 -0.13 -0.12 2.20 6 49.39 20460.6 20464.6 9381.4 -0.10 -0.07 2.18 7 61.73 25580.3 25580.1 11906.7 -0.02 -0.02 2.15 8 82.31 34119.9 34122.0 15598.3 0.06 0.07 2.19 9 102.89 42680.8 42680.0 19818.6 0.16 0.16 2.15 10 123.46 51255.4 51248.3 24038.2 0.26 0.25 2.13 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3752 to 109653 e- Average count rate measured: 414 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.14 +/- 0.01 e-/ADU RON = 5.33 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.116 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay