*** MIDAS logfile (vers. 99NOV) *** Su, 25 Feb 2001 15:58:06 Page 1 *** MIDAS logfile (vers. 99NOV) *** Su, 25 Feb 2001 15:58:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010225 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.89 stddev = 2.47 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010225 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1757.4 1757.1 813.5 0.08 0.10 2.16 2 8.23 3465.4 3463.7 1589.4 -0.02 -0.05 2.18 3 16.47 6883.2 6883.4 3127.7 -0.17 -0.16 2.20 4 24.70 10303.7 10304.2 4741.5 -0.15 -0.15 2.17 5 37.04 15435.3 15438.3 7103.9 -0.12 -0.10 2.17 6 49.39 20578.1 20577.6 9526.9 -0.07 -0.08 2.16 7 61.73 25719.3 25721.8 11698.0 -0.03 -0.03 2.20 8 82.31 34315.8 34314.1 15909.5 0.07 0.06 2.16 9 102.89 42923.7 42922.5 19827.9 0.16 0.15 2.16 10 123.46 51551.5 51546.3 23663.8 0.27 0.25 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3814 to 111896 e- Average count rate measured: 416 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.01 e-/ADU RON = 5.36 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.099 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay