*** MIDAS logfile (vers. 99NOV) *** Tu, 30 Jan 2001 17:10:05 Page 1 *** MIDAS logfile (vers. 99NOV) *** Tu, 30 Jan 2001 17:10:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.50 stddev = 2.45 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1786.7 1785.0 922.7 0.11 0.14 1.94 2 8.23 3514.5 3511.1 1599.8 -0.08 -0.12 2.20 3 16.47 6978.6 6977.1 3183.7 -0.18 -0.18 2.19 4 24.70 10443.5 10445.0 4751.9 -0.16 -0.14 2.20 5 37.04 15644.8 15642.3 7254.7 -0.11 -0.12 2.16 6 49.39 20851.9 20855.4 9499.1 -0.08 -0.06 2.20 7 61.73 26066.3 26067.0 11846.7 -0.02 -0.02 2.20 8 82.31 34770.7 34773.2 15852.8 0.07 0.07 2.19 9 102.89 43500.4 43495.2 20138.7 0.18 0.16 2.16 10 123.46 52235.5 52233.2 23986.5 0.27 0.26 2.18 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3889 to 113733 e- Average count rate measured: 422 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.18 +/- 0.01 e-/ADU RON = 5.34 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.111 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.27/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay