*** MIDAS logfile (vers. 99NOV) *** Tu, 09 Jan 2001 16:12:29 Page 1 *** MIDAS logfile (vers. 99NOV) *** Tu, 09 Jan 2001 16:13:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010109 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 578.96 stddev = 2.49 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_36_010109 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.12 1531.0 1529.2 718.3 0.14 0.11 2.13 2 8.23 3017.7 3018.0 1378.4 -0.17 -0.11 2.19 3 16.47 6011.1 6008.2 2775.3 -0.11 -0.14 2.17 4 24.70 8995.0 8996.1 4144.7 -0.16 -0.14 2.17 5 37.04 13481.4 13482.1 6264.7 -0.10 -0.09 2.15 6 49.39 17973.1 17972.8 8424.8 -0.06 -0.06 2.13 7 61.73 22467.6 22465.1 10303.8 -0.01 -0.02 2.18 8 82.31 29970.0 29968.1 13810.3 0.07 0.06 2.17 9 102.89 37494.0 37491.7 17065.4 0.17 0.16 2.20 10 123.46 45014.8 45010.7 20922.2 0.24 0.23 2.15 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3320 to 97671 e- Average count rate measured: 363 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.17 +/- 0.01 e-/ADU RON = 5.40 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.084 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay