*** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 23 Mar 2002 16:34:47 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Sa, 23 Mar 2002 16:35:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 020323 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.63 stddev = 2.25 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_020323 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1800.5 1800.6 757.9 0.31 0.30 2.38 2 10.00 3564.9 3565.7 1478.2 -0.22 -0.21 2.41 3 20.00 7095.7 7095.3 3003.6 -0.46 -0.47 2.36 4 30.00 10637.2 10637.7 4500.5 -0.44 -0.44 2.36 5 45.00 15962.5 15964.5 6690.1 -0.35 -0.33 2.39 6 60.00 21307.2 21308.7 9006.8 -0.21 -0.20 2.37 7 75.00 26671.5 26670.2 11262.4 -0.05 -0.05 2.37 8 100.00 35649.1 35650.6 15147.8 0.21 0.22 2.35 9 125.00 44671.0 44666.8 19233.6 0.47 0.46 2.32 10 150.00 53746.7 53739.4 23009.8 0.74 0.73 2.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4194 to 125173 e- Average count rate measured: 356 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.33 +/- 0.01 e-/ADU RON = 5.24 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.048 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.73/-0.46 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay