*** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 12 Feb 2002 14:27:45 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Tu, 12 Feb 2002 14:28:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 020212 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.89 stddev = 2.26 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_020212 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1833.9 1834.7 763.5 0.30 0.33 2.40 2 10.00 3620.3 3617.2 1488.6 -0.21 -0.30 2.43 3 20.00 7194.5 7196.1 3002.5 -0.45 -0.43 2.40 4 30.00 10778.7 10778.2 4472.1 -0.44 -0.44 2.41 5 45.00 16169.9 16170.1 6751.8 -0.34 -0.33 2.39 6 60.00 21578.3 21577.0 8968.3 -0.21 -0.21 2.41 7 75.00 27007.1 27010.5 11513.6 -0.06 -0.04 2.35 8 100.00 36096.4 36098.2 15313.6 0.21 0.22 2.36 9 125.00 45230.8 45225.5 19411.7 0.47 0.47 2.33 10 150.00 54417.5 54409.6 23419.9 0.74 0.73 2.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4255 to 126230 e- Average count rate measured: 360 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.32 +/- 0.01 e-/ADU RON = 5.24 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.080 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.74/-0.45 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay