*** MIDAS logfile (vers. 01FEBpl1.1) *** We, 06 Feb 2002 18:23:50 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** We, 06 Feb 2002 18:24:31 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 020206 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.49 stddev = 2.25 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_020206 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1833.8 1834.2 782.6 0.32 0.32 2.34 2 10.00 3622.8 3622.6 1540.2 -0.26 -0.27 2.35 3 20.00 7207.1 7207.4 2959.4 -0.46 -0.46 2.44 4 30.00 10801.9 10801.1 4505.6 -0.43 -0.44 2.40 5 45.00 16206.5 16207.4 6796.7 -0.34 -0.33 2.38 6 60.00 21628.2 21628.9 9179.1 -0.21 -0.21 2.36 7 75.00 27073.1 27071.2 11318.6 -0.05 -0.05 2.39 8 100.00 36183.5 36183.9 15358.7 0.21 0.22 2.36 9 125.00 45346.4 45343.6 19344.0 0.48 0.48 2.34 10 150.00 54549.3 54545.1 23597.6 0.74 0.73 2.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4258 to 126638 e- Average count rate measured: 361 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.32 +/- 0.01 e-/ADU RON = 5.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.066 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.74/-0.46 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay