*** MIDAS logfile (vers. 01FEBpl1.1) *** We, 30 Jan 2002 14:45:48 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** We, 30 Jan 2002 14:46:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 020130 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.84 stddev = 2.26 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_020130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1842.2 1843.5 767.3 0.32 0.30 2.40 2 10.00 3645.2 3647.8 1506.4 -0.25 -0.23 2.42 3 20.00 7257.5 7259.1 3023.2 -0.45 -0.45 2.40 4 30.00 10878.8 10877.7 4538.3 -0.44 -0.46 2.40 5 45.00 16324.6 16325.6 6792.0 -0.34 -0.35 2.40 6 60.00 21788.7 21791.3 9154.2 -0.21 -0.21 2.38 7 75.00 27273.7 27277.1 11518.7 -0.06 -0.05 2.37 8 100.00 36456.6 36457.8 15586.2 0.21 0.21 2.34 9 125.00 45685.8 45690.6 19664.8 0.47 0.48 2.32 10 150.00 54965.1 54963.6 23682.2 0.74 0.74 2.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4265 to 127214 e- Average count rate measured: 364 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.31 +/- 0.01 e-/ADU RON = 5.23 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.052 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.74/-0.46 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay