*** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 06 Jan 2002 16:10:50 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 06 Jan 2002 16:11:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 020106 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 584.94 stddev = 2.27 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_020106 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1853.5 1855.7 770.9 0.33 0.32 2.40 2 10.00 3663.2 3665.3 1508.1 -0.26 -0.25 2.43 3 20.00 7289.7 7289.8 3044.6 -0.45 -0.47 2.39 4 30.00 10923.5 10923.6 4591.2 -0.45 -0.46 2.38 5 45.00 16389.8 16391.1 6752.5 -0.36 -0.35 2.43 6 60.00 21879.5 21879.1 9270.6 -0.20 -0.20 2.36 7 75.00 27385.5 27385.8 11641.7 -0.05 -0.05 2.35 8 100.00 36606.5 36605.8 15547.4 0.22 0.23 2.35 9 125.00 45871.5 45872.7 19585.3 0.48 0.49 2.34 10 150.00 55186.5 55186.0 23499.1 0.75 0.76 2.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4338 to 129095 e- Average count rate measured: 365 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.34 +/- 0.01 e-/ADU RON = 5.30 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.064 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.75/-0.46 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay