*** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 30 Dec 2001 15:45:04 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Su, 30 Dec 2001 15:45:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011230 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.87 stddev = 2.26 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_011230 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1881.0 1880.4 797.6 0.32 0.30 2.36 2 10.00 3719.0 3720.3 1572.2 -0.27 -0.23 2.37 3 20.00 7403.6 7402.5 3056.2 -0.45 -0.46 2.42 4 30.00 11095.8 11094.5 4884.7 -0.44 -0.45 2.27 5 45.00 16649.1 16651.1 6939.7 -0.34 -0.33 2.40 6 60.00 22219.2 22220.4 9397.2 -0.22 -0.22 2.36 7 75.00 27816.4 27814.3 11867.7 -0.05 -0.06 2.34 8 100.00 37179.9 37180.8 15716.0 0.22 0.22 2.37 9 125.00 46588.6 46593.9 20022.9 0.47 0.48 2.33 10 150.00 56060.2 56052.7 24055.1 0.76 0.74 2.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4384 to 130679 e- Average count rate measured: 371 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.33 +/- 0.01 e-/ADU RON = 5.28 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.056 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.75/-0.45 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay