*** MIDAS logfile (vers. 01FEBpl1.1) *** Fr, 30 Nov 2001 13:57:15 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Fr, 30 Nov 2001 13:57:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 011130 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 583.94 stddev = 2.25 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_011130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1894.1 1896.7 787.0 0.32 0.35 2.41 2 10.00 3734.9 3734.0 1546.9 -0.24 -0.31 2.41 3 20.00 7421.1 7422.6 3111.5 -0.46 -0.46 2.39 4 30.00 11116.5 11118.5 4662.2 -0.45 -0.44 2.38 5 45.00 16676.7 16678.3 6960.3 -0.34 -0.34 2.40 6 60.00 22253.8 22256.3 9385.0 -0.21 -0.20 2.37 7 75.00 27850.8 27853.8 11835.6 -0.06 -0.05 2.35 8 100.00 37225.4 37224.7 15956.6 0.21 0.21 2.33 9 125.00 46649.9 46652.8 20114.3 0.48 0.49 2.32 10 150.00 56127.5 56121.1 24223.5 0.76 0.75 2.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4379 to 129674 e- Average count rate measured: 371 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.31 +/- 0.01 e-/ADU RON = 5.21 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.090 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.76/-0.46 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay