*** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 16 Aug 2001 13:47:08 Page 1 *** MIDAS logfile (vers. 01FEBpl1.1) *** Th, 16 Aug 2001 13:47:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010816 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 584.08 stddev = 2.28 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_010816 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1964.6 1964.0 818.6 0.34 0.33 2.40 2 10.00 3858.5 3858.9 1642.7 -0.27 -0.25 2.35 3 20.00 7654.0 7655.3 3174.7 -0.48 -0.46 2.41 4 30.00 11457.9 11458.1 4795.3 -0.48 -0.47 2.39 5 45.00 17183.2 17183.5 7163.2 -0.36 -0.36 2.40 6 60.00 22928.4 22927.7 9693.1 -0.22 -0.22 2.37 7 75.00 28697.3 28697.9 12141.5 -0.05 -0.05 2.36 8 100.00 38354.2 38351.4 16473.8 0.23 0.22 2.33 9 125.00 48072.5 48068.9 20964.4 0.53 0.52 2.29 10 150.00 57811.7 57805.1 25065.0 0.76 0.75 2.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4508 to 132669 e- Average count rate measured: 382 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.29 +/- 0.01 e-/ADU RON = 5.23 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.122 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.75/-0.48 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay