*** MIDAS logfile (vers. 99NOV) *** We, 09 May 2001 00:09:54 Page 1 *** MIDAS logfile (vers. 99NOV) *** We, 09 May 2001 00:10:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010509 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 584.67 stddev = 2.26 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_010509 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1952.3 1952.6 827.6 0.36 0.36 2.36 2 10.00 3855.5 3855.4 1568.7 -0.31 -0.32 2.46 3 20.00 7672.4 7675.9 3208.4 -0.51 -0.47 2.39 4 30.00 11501.8 11503.1 4802.4 -0.47 -0.46 2.40 5 45.00 17259.3 17259.3 7209.8 -0.36 -0.36 2.39 6 60.00 23038.3 23038.3 9860.7 -0.22 -0.22 2.34 7 75.00 28843.4 28840.1 12041.3 -0.04 -0.05 2.40 8 100.00 38556.8 38551.3 16728.8 0.24 0.22 2.30 9 125.00 48333.1 48324.0 20837.6 0.53 0.52 2.32 10 150.00 58137.9 58130.2 24956.3 0.78 0.77 2.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4518 to 134528 e- Average count rate measured: 384 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.31 +/- 0.01 e-/ADU RON = 5.24 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.060 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.78/-0.49 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay