*** MIDAS logfile (vers. 99NOV) *** Tu, 08 May 2001 23:27:53 Page 1 *** MIDAS logfile (vers. 99NOV) *** Tu, 08 May 2001 23:28:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010508 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 584.20 stddev = 2.27 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_010508 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1947.7 1951.0 814.2 0.30 0.36 2.39 2 10.00 3857.0 3854.7 1600.1 -0.19 -0.30 2.41 3 20.00 7673.8 7673.9 3214.5 -0.47 -0.47 2.39 4 30.00 11501.5 11500.5 4805.9 -0.46 -0.47 2.39 5 45.00 17257.4 17257.9 7299.5 -0.38 -0.36 2.36 6 60.00 23040.9 23040.2 9781.2 -0.22 -0.21 2.36 7 75.00 28841.2 28840.4 12239.3 -0.06 -0.05 2.36 8 100.00 38551.2 38550.9 16647.1 0.21 0.22 2.32 9 125.00 48328.5 48324.1 21054.8 0.51 0.52 2.30 10 150.00 58133.9 58124.7 25186.1 0.76 0.76 2.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 4474 to 133411 e- Average count rate measured: 384 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.30 +/- 0.01 e-/ADU RON = 5.21 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.053 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.76/-0.47 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay