*** MIDAS logfile (vers. 99NOV) *** Tu, 09 Jan 2001 15:33:57 Page 1 *** MIDAS logfile (vers. 99NOV) *** Tu, 09 Jan 2001 15:34:33 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#36 TAKEN ON 010109 IN fastcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 584.29 stddev = 2.25 There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_fastcd_36_010109 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 5.00 1668.7 1669.5 696.1 0.27 0.27 2.40 2 10.00 3309.7 3310.3 1345.6 -0.19 -0.19 2.46 3 20.00 6590.5 6591.0 2725.0 -0.43 -0.43 2.42 4 30.00 9883.8 9884.7 4122.9 -0.39 -0.38 2.40 5 45.00 14831.3 14831.9 6243.4 -0.31 -0.31 2.38 6 60.00 19793.9 19795.6 8196.9 -0.19 -0.18 2.41 7 75.00 24772.3 24771.6 10452.3 -0.06 -0.06 2.37 8 100.00 33106.5 33104.6 13920.8 0.19 0.18 2.38 9 125.00 41481.7 41479.7 17519.3 0.43 0.43 2.37 10 150.00 49898.2 49894.3 21350.9 0.68 0.68 2.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 3914 to 117018 e- Average count rate measured: 330 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 2.35 +/- 0.01 e-/ADU RON = 5.28 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.039 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.68/-0.43 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay