*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 15:02:47 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 15:03:44 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 061018 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 291.93 stddev = 3.27 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rC2_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1709.1 1709.8 1407.4 -0.08 -0.07 1.21 2 2.12 3363.8 3363.5 2926.0 0.17 0.15 1.15 3 4.25 6641.7 6640.7 5462.1 -0.09 -0.10 1.22 4 6.38 9933.8 9934.0 8057.9 -0.04 -0.03 1.23 5 9.57 14872.1 14871.1 12246.2 0.00 0.00 1.21 6 12.76 19824.3 19823.2 16032.0 0.09 0.09 1.24 7 15.95 24758.5 24759.6 19996.3 0.08 0.08 1.24 8 21.27 32981.7 32981.4 26448.1 0.05 0.05 1.25 9 26.59 41180.0 41181.9 32522.0 -0.02 -0.01 1.27 10 31.91 49330.8 49331.6 38655.1 -0.16 -0.16 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2183 to 62987 e- Average count rate measured: 1546 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.01 e-/ADU RON = 4.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.048 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay