*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 15:57:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 15:58:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060121 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 292.11 stddev = 3.38 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rC2_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 930.3 930.1 771.3 0.02 0.02 1.21 2 2.00 1821.6 1821.3 1520.1 -0.04 -0.06 1.20 3 4.00 3605.0 3604.6 3084.9 -0.04 -0.04 1.17 4 6.00 5387.3 5387.6 4623.0 -0.04 -0.03 1.17 5 9.00 8062.3 8061.9 6954.6 -0.03 -0.03 1.16 6 12.00 10738.5 10739.1 9283.6 -0.01 -0.00 1.16 7 15.00 13417.3 13416.0 10959.7 0.02 0.02 1.22 8 20.00 17877.6 17878.3 14560.5 0.03 0.03 1.23 9 25.00 22339.2 22339.9 18081.3 0.04 0.04 1.24 10 30.00 26800.5 26802.0 21486.5 0.04 0.05 1.25 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1164 to 33536 e- Average count rate measured: 892 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 4.22 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.046 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.05/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay