*** MIDAS logfile (05SEPpl1.2) *** Sa, 04 Nov 2006 16:52:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 04 Nov 2006 16:53:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA2 TAKEN ON 061104 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 280.66 stddev = 3.02 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rA2_061104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1930.2 1930.9 1554.1 -0.22 -0.21 1.24 2 2.12 3721.3 3720.9 2998.8 0.26 0.25 1.24 3 4.25 7268.6 7267.5 5802.8 0.08 0.07 1.25 4 6.38 10829.6 10829.4 8850.2 0.13 0.13 1.22 5 9.57 16164.1 16162.2 15612.2 0.13 0.12 1.04 6 12.76 21505.6 21504.1 17076.3 0.16 0.16 1.26 7 15.95 26823.4 26823.6 21016.7 0.09 0.10 1.28 8 21.27 35668.8 35670.6 27988.9 -0.03 -0.02 1.27 9 26.59 44473.4 44481.8 33956.4 -0.19 -0.17 1.31 10 31.91 53214.8 53214.8 40393.7 -0.42 -0.42 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2564 to 70677 e- Average count rate measured: 1669 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.03 e-/ADU RON = 4.01 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.100 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.26/-0.42 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay