*** MIDAS logfile (05SEPpl1.2) *** Th, 07 Dec 2006 14:22:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 07 Dec 2006 14:23:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 061207 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 270.01 stddev = 3.09 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rA1_061207 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1796.8 1796.7 1487.6 -0.10 -0.11 1.21 2 8.51 3557.3 3557.5 2818.5 0.06 0.07 1.26 3 17.02 7077.5 7077.1 5661.7 0.14 0.13 1.25 4 25.53 10595.6 10595.5 8263.0 0.14 0.14 1.28 5 38.30 15875.6 15874.8 12214.8 0.16 0.16 1.30 6 51.06 21143.1 21143.1 15867.3 0.12 0.12 1.33 7 63.83 26412.3 26412.0 19537.3 0.09 0.10 1.35 8 85.10 35157.2 35156.3 25164.0 -0.03 -0.03 1.40 9 106.38 43873.6 43870.9 30431.2 -0.18 -0.18 1.44 10 127.66 52518.7 52519.9 35317.8 -0.41 -0.40 1.49 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2673 to 78124 e- Average count rate measured: 413 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.49 +/- 0.03 e-/ADU RON = 4.60 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.106 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.40 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay