*** MIDAS logfile (05SEPpl1.2) *** We, 18 Oct 2006 13:56:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 18 Oct 2006 13:57:37 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 061018 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 270.53 stddev = 2.86 FIERA temperature = 287.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rA1_061018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1710.8 1710.9 1405.8 -0.09 -0.09 1.22 2 8.51 3401.1 3401.3 2696.9 0.06 0.07 1.26 3 17.02 6779.0 6779.3 5349.6 0.12 0.12 1.27 4 25.53 10157.1 10156.4 7890.8 0.14 0.13 1.29 5 38.29 15223.3 15223.7 11698.3 0.14 0.15 1.30 6 51.06 20281.3 20282.0 15295.3 0.11 0.11 1.33 7 63.83 25338.0 25338.8 18738.8 0.08 0.08 1.35 8 85.10 33740.6 33740.2 24254.6 -0.02 -0.03 1.39 9 106.38 42107.6 42106.6 29488.9 -0.17 -0.17 1.43 10 127.66 50425.4 50426.6 34185.7 -0.37 -0.36 1.48 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2520 to 74264 e- Average count rate measured: 396 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.47 +/- 0.03 e-/ADU RON = 4.21 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.070 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.36 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay