*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Jul 2006 16:56:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Jul 2006 16:58:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060720 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 230.42 stddev = 2.93 FIERA temperature = 288.4 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rA1_060720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.26 1763.0 1765.6 1395.9 -0.11 -0.04 1.26 2 8.51 3491.2 3490.0 2750.1 0.12 0.00 1.27 3 17.02 6939.1 6940.4 5571.2 0.08 0.06 1.25 4 25.53 10388.8 10390.6 7984.2 0.07 0.08 1.30 5 38.30 15568.8 15570.3 11779.4 0.12 0.13 1.32 6 51.07 20740.4 20741.0 17028.4 0.10 0.10 1.22 7 63.83 25913.2 25913.4 19306.8 0.09 0.09 1.34 8 85.11 34502.3 34506.7 24532.6 -0.01 0.01 1.41 9 106.38 43062.5 43062.6 29777.6 -0.14 -0.13 1.45 10 127.66 51573.3 51572.5 34341.2 -0.32 -0.31 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2641 to 77196 e- Average count rate measured: 405 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.50 +/- 0.04 e-/ADU RON = 4.38 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.105 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.31 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay