*** MIDAS logfile (05SEPpl1.2) *** Th, 06 Apr 2006 13:38:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 06 Apr 2006 13:39:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rA1 TAKEN ON 060406 IN slowcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 272.53 stddev = 3.01 FIERA temperature = 289.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_slowcd_rA1_060406 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1874.5 1874.6 1501.9 -0.12 -0.12 1.25 2 8.51 3713.3 3713.5 2935.2 0.08 0.09 1.27 3 17.02 7387.6 7387.2 5789.9 0.15 0.14 1.28 4 25.53 11061.8 11061.7 8605.0 0.17 0.16 1.29 5 38.30 16571.9 16571.8 12651.2 0.17 0.17 1.31 6 51.06 22073.0 22073.9 16528.3 0.14 0.14 1.34 7 63.83 27567.4 27568.9 20276.6 0.09 0.09 1.36 8 85.10 36698.7 36698.6 26042.9 -0.03 -0.03 1.41 9 106.38 45778.9 45778.4 31329.9 -0.21 -0.22 1.46 10 127.66 54807.7 54807.7 36414.3 -0.42 -0.43 1.51 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2822 to 82502 e- Average count rate measured: 431 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.51 +/- 0.03 e-/ADU RON = 4.54 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.103 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.43 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay