*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 16:15:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 16:15:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.69 stddev = 5.42 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1312.4 1312.4 1056.1 -0.00 -0.01 1.24 2 2.12 2632.5 2632.1 2091.7 0.14 0.13 1.26 3 4.25 5249.1 5248.7 4154.0 -0.19 -0.19 1.26 4 6.38 7879.7 7879.4 6289.6 -0.13 -0.13 1.25 5 9.57 11824.0 11823.9 9236.8 -0.09 -0.09 1.28 6 12.76 15782.7 15784.3 12414.3 0.01 0.02 1.27 7 15.95 19733.1 19735.3 15281.9 0.04 0.05 1.29 8 21.27 26325.5 26326.3 20414.9 0.09 0.09 1.29 9 26.59 32909.1 32908.3 24986.9 0.09 0.09 1.32 10 31.91 39467.5 39467.4 30116.6 0.03 0.04 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1727 to 51947 e- Average count rate measured: 1236 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.13 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay