*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Jan 2008 15:27:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Jan 2008 15:27:23 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 080110 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 287.36 stddev = 5.83 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_080110 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1377.1 1377.1 1114.3 -0.03 -0.01 1.24 2 2.12 2736.6 2736.6 2168.8 0.17 0.17 1.26 3 4.25 5429.6 5428.4 4289.9 -0.18 -0.19 1.27 4 6.38 8137.5 8138.0 6406.2 -0.12 -0.12 1.27 5 9.57 12197.9 12196.6 10044.3 -0.08 -0.09 1.21 6 12.76 16273.5 16273.5 12763.2 0.02 0.02 1.28 7 15.95 20338.3 20340.4 16019.1 0.04 0.05 1.27 8 21.27 27121.9 27121.6 21101.5 0.08 0.08 1.29 9 26.59 33897.8 33897.1 26473.0 0.08 0.08 1.28 10 31.91 40643.5 40642.9 30955.0 0.02 0.02 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1796 to 53016 e- Average count rate measured: 1273 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 7.61 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.024 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay