*** MIDAS logfile (05SEPpl1.2) *** Sa, 08 Dec 2007 16:20:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 08 Dec 2007 16:20:22 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 071208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.28 stddev = 5.66 FIERA temperature = 286.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_071208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1388.6 1389.3 1481.9 -0.00 -0.03 0.94 2 2.12 2757.2 2758.1 2189.3 0.20 0.17 1.26 3 4.25 5469.6 5468.3 4930.5 -0.16 -0.18 1.11 4 6.38 8195.3 8195.9 6406.6 -0.12 -0.12 1.28 5 9.57 12283.3 12284.3 9665.0 -0.09 -0.08 1.27 6 12.76 16387.0 16388.1 13185.3 0.02 0.03 1.24 7 15.95 20479.6 20480.3 16051.1 0.03 0.05 1.28 8 21.27 27308.9 27307.8 21212.7 0.07 0.08 1.29 9 26.59 34128.1 34128.0 26233.7 0.07 0.08 1.30 10 31.91 40917.6 40917.8 31233.3 -0.01 0.01 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1836 to 54073 e- Average count rate measured: 1281 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.47 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.024 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay