*** MIDAS logfile (05SEPpl1.2) *** Sa, 03 Nov 2007 15:56:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 03 Nov 2007 15:57:12 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 071103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 333.31 stddev = 5.30 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_071103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1432.3 1433.1 1182.5 -0.09 -0.14 1.21 2 2.12 2836.7 2836.4 2312.3 0.30 0.27 1.23 3 4.25 5608.4 5609.6 4491.5 -0.08 -0.08 1.25 4 6.38 8397.0 8397.1 6735.1 -0.04 -0.04 1.25 5 9.57 12577.6 12576.4 10052.5 -0.02 -0.01 1.25 6 12.76 16765.0 16766.5 13342.5 0.04 0.05 1.26 7 15.95 20948.6 20948.1 16708.8 0.04 0.05 1.25 8 21.27 27912.0 27912.6 22635.7 0.02 0.04 1.23 9 26.59 34863.9 34864.7 27063.3 -0.02 -0.01 1.29 10 31.91 41779.2 41779.1 31874.4 -0.14 -0.13 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1859 to 54206 e- Average count rate measured: 1309 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.02 e-/ADU RON = 6.87 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.036 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.28/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay