*** MIDAS logfile (05SEPpl1.2) *** Th, 18 Oct 2007 17:11:25 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 18 Oct 2007 17:11:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 071018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 335.01 stddev = 5.29 FIERA temperature = 286.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_071018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1412.4 1412.8 1155.3 -0.10 -0.10 1.22 2 2.12 2802.6 2802.2 2289.4 0.26 0.24 1.22 3 4.25 5552.3 5551.7 4461.7 -0.09 -0.09 1.24 4 6.38 8316.9 8317.1 6684.8 -0.04 -0.04 1.24 5 9.57 12458.8 12460.0 9829.7 -0.03 -0.02 1.27 6 12.76 16614.7 16615.8 13248.4 0.05 0.06 1.25 7 15.95 20759.7 20760.0 16665.3 0.05 0.05 1.25 8 21.27 27664.6 27664.5 21693.0 0.03 0.03 1.28 9 26.59 34558.8 34560.8 27096.9 -0.01 -0.00 1.28 10 31.91 41415.4 41416.7 32069.3 -0.13 -0.12 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1820 to 53364 e- Average count rate measured: 1298 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 6.82 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.030 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay