*** MIDAS logfile (05SEPpl1.2) *** Su, 14 Oct 2007 16:12:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 Oct 2007 16:12:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 071014 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 343.69 stddev = 5.61 FIERA temperature = 285.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_071014 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1439.8 1439.8 1171.4 -0.12 -0.08 1.23 2 2.12 2847.1 2846.2 2303.5 0.24 0.25 1.24 3 4.25 5630.1 5629.7 4508.3 -0.09 -0.08 1.25 4 6.38 8428.2 8427.1 6759.5 -0.04 -0.05 1.25 5 9.57 12621.6 12619.8 10122.6 -0.02 -0.03 1.25 6 12.76 16826.9 16826.4 13377.0 0.06 0.05 1.26 7 15.95 21020.3 21021.1 16519.9 0.05 0.05 1.27 8 21.27 28009.9 28010.2 21942.2 0.04 0.03 1.28 9 26.59 34986.7 34988.0 26977.7 -0.00 -0.01 1.30 10 31.91 41923.1 41921.9 32461.0 -0.12 -0.14 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1868 to 54401 e- Average count rate measured: 1314 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 7.29 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.037 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay