*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Sep 2007 18:25:10 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Sep 2007 18:25:25 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.29 stddev = 5.72 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1373.7 1375.0 1221.2 -0.08 -0.10 1.12 2 2.12 2752.5 2752.7 2218.5 0.23 0.21 1.24 3 4.25 5484.4 5485.3 4425.6 -0.10 -0.10 1.24 4 6.38 8231.3 8230.7 6542.8 -0.04 -0.05 1.26 5 9.57 12346.7 12345.9 9962.4 -0.03 -0.03 1.24 6 12.76 16474.3 16475.5 13175.9 0.05 0.06 1.25 7 15.95 20592.1 20589.7 16317.3 0.05 0.05 1.26 8 21.27 27452.3 27452.6 21869.0 0.04 0.04 1.26 9 26.59 34302.1 34302.2 26438.8 -0.01 0.01 1.30 10 31.91 41114.7 41115.7 31857.7 -0.12 -0.10 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1779 to 53230 e- Average count rate measured: 1290 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 7.41 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay