*** MIDAS logfile (05SEPpl1.2) *** Sa, 15 Sep 2007 17:26:07 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 15 Sep 2007 17:26:21 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070915 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 343.35 stddev = 5.60 FIERA temperature = 282.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070915 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1347.1 1347.7 1097.1 -0.09 -0.07 1.23 2 2.12 2722.8 2722.8 2207.3 0.20 0.19 1.23 3 4.25 5449.0 5449.5 4378.5 -0.10 -0.10 1.24 4 6.38 8188.1 8188.4 10622.0 -0.05 -0.05 0.77 5 9.57 12293.8 12293.4 9822.5 -0.02 -0.03 1.25 6 12.76 16411.6 16412.9 13079.5 0.05 0.05 1.25 7 15.95 20520.7 20520.4 16187.2 0.06 0.06 1.27 8 21.27 27364.3 27365.6 21603.1 0.05 0.05 1.27 9 26.59 34197.9 34197.7 26617.5 0.01 0.00 1.28 10 31.91 40995.3 40997.1 31653.3 -0.10 -0.10 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1792 to 54528 e- Average count rate measured: 1286 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.06 e-/ADU RON = 7.45 +/- 0.33 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.011 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay