*** MIDAS logfile (05SEPpl1.2) *** We, 05 Sep 2007 17:03:28 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 05 Sep 2007 17:03:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070905 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 339.53 stddev = 5.49 FIERA temperature = 283.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070905 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1380.4 1379.3 1131.8 -0.11 -0.09 1.22 2 2.12 2767.5 2767.3 2239.3 0.19 0.23 1.24 3 4.25 5513.9 5513.0 4404.7 -0.10 -0.11 1.25 4 6.38 8275.0 8275.2 6658.7 -0.03 -0.04 1.24 5 9.57 12413.0 12411.4 9847.1 -0.02 -0.02 1.26 6 12.76 16560.4 16561.3 13140.5 0.06 0.05 1.26 7 15.95 20701.1 20700.7 16349.2 0.06 0.05 1.27 8 21.27 27597.2 27597.5 21740.0 0.05 0.04 1.27 9 26.59 34480.9 34480.3 26970.5 0.01 -0.00 1.28 10 31.91 41329.6 41331.5 31897.0 -0.10 -0.11 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1782 to 53387 e- Average count rate measured: 1296 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 7.09 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.006 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay