*** MIDAS logfile (05SEPpl1.2) *** Mo, 27 Aug 2007 15:53:47 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 27 Aug 2007 15:54:01 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070827 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 335.07 stddev = 5.30 FIERA temperature = 286.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070827 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1355.3 1356.5 1126.2 -0.09 -0.05 1.20 2 2.12 2738.6 2738.6 2201.4 0.24 0.24 1.24 3 4.25 5473.9 5474.6 4367.8 -0.11 -0.11 1.25 4 6.38 8224.9 8225.5 6614.8 -0.05 -0.05 1.24 5 9.57 12348.8 12349.0 9830.6 -0.03 -0.03 1.26 6 12.76 16484.7 16484.8 15005.9 0.06 0.05 1.10 7 15.95 20608.9 20607.2 16454.0 0.06 0.04 1.25 8 21.27 27481.0 27480.4 21611.9 0.04 0.03 1.27 9 26.59 34339.2 34341.1 26421.4 -0.00 -0.01 1.30 10 31.91 41166.0 41166.1 31753.6 -0.11 -0.12 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1763 to 53536 e- Average count rate measured: 1292 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.03 e-/ADU RON = 6.90 +/- 0.16 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay