*** MIDAS logfile (05SEPpl1.2) *** Sa, 28 Apr 2007 18:13:24 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 28 Apr 2007 18:13:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070428 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.39 stddev = 6.77 FIERA temperature = 282.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070428 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1392.8 1393.0 1213.5 -0.09 -0.10 1.15 2 2.12 2812.1 2811.1 2301.5 0.21 0.19 1.22 3 4.25 5623.2 5622.2 4567.1 -0.09 -0.10 1.23 4 6.38 8448.2 8447.0 6752.0 -0.05 -0.05 1.25 5 9.57 12681.9 12681.7 10207.2 -0.02 -0.02 1.24 6 12.76 16930.3 16928.0 13418.4 0.06 0.06 1.26 7 15.95 21164.9 21164.8 17176.3 0.05 0.06 1.23 8 21.27 28223.9 28224.9 22405.4 0.05 0.05 1.26 9 26.59 35267.8 35269.8 27573.9 -0.00 0.01 1.28 10 31.91 42275.2 42277.7 32425.1 -0.12 -0.10 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1805 to 54782 e- Average count rate measured: 1327 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 8.77 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.009 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay