*** MIDAS logfile (05SEPpl1.2) *** Su, 25 Mar 2007 15:44:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 25 Mar 2007 15:44:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070325 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.08 stddev = 5.68 FIERA temperature = 282.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070325 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1396.5 1396.1 1139.7 -0.12 -0.09 1.23 2 2.12 2828.4 2827.4 2339.4 0.22 0.25 1.21 3 4.25 5657.9 5659.2 4614.1 -0.11 -0.09 1.23 4 6.38 8504.0 8503.4 6851.1 -0.04 -0.05 1.24 5 9.57 12769.5 12768.5 10198.3 -0.02 -0.04 1.25 6 12.76 17047.6 17048.5 13459.5 0.06 0.05 1.27 7 15.95 21316.0 21315.3 16737.8 0.06 0.05 1.27 8 21.27 28425.0 28425.3 22728.9 0.05 0.04 1.25 9 26.59 35521.3 35520.0 27756.6 0.01 -0.01 1.28 10 31.91 42576.6 42580.1 32710.1 -0.12 -0.13 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1808 to 55119 e- Average count rate measured: 1336 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 7.35 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.014 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay