*** MIDAS logfile (05SEPpl1.2) *** Mo, 12 Mar 2007 17:08:36 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 12 Mar 2007 17:08:51 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070312 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.16 stddev = 5.58 FIERA temperature = 283.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070312 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1433.2 1432.9 1183.2 -0.11 -0.13 1.21 2 2.13 2876.4 2875.2 2321.5 0.24 0.22 1.24 3 4.25 5728.5 5728.3 4589.9 -0.09 -0.10 1.25 4 6.38 8597.9 8596.9 7465.7 -0.04 -0.04 1.15 5 9.57 12895.6 12896.9 10303.1 -0.03 -0.01 1.25 6 12.76 17208.1 17208.2 13640.3 0.06 0.06 1.26 7 15.95 21508.3 21507.7 17147.7 0.06 0.06 1.25 8 21.27 28676.0 28674.5 22568.3 0.05 0.05 1.27 9 26.59 35824.5 35825.7 27704.9 -0.01 0.01 1.29 10 31.91 42934.2 42937.4 32972.9 -0.13 -0.12 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1871 to 56053 e- Average count rate measured: 1347 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 7.29 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.005 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.23/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay