*** MIDAS logfile (05SEPpl1.2) *** We, 03 Jan 2007 16:36:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Jan 2007 16:36:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 070103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 352.94 stddev = 5.57 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_070103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1429.3 1427.2 1174.7 -0.10 -0.10 1.22 2 2.12 2886.9 2885.8 2330.1 0.24 0.26 1.24 3 4.25 5769.2 5767.5 4614.0 -0.09 -0.10 1.25 4 6.38 8667.1 8667.0 6913.6 -0.04 -0.04 1.25 5 9.57 13010.7 13010.0 10404.6 -0.03 -0.03 1.25 6 12.76 17368.7 17369.9 13878.7 0.06 0.06 1.25 7 15.95 21714.4 21715.4 17142.9 0.06 0.06 1.27 8 21.27 28957.1 28958.5 23006.2 0.05 0.04 1.26 9 26.59 36180.1 36181.0 27847.7 -0.01 -0.01 1.30 10 31.91 43367.8 43366.8 33957.4 -0.13 -0.14 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1837 to 55772 e- Average count rate measured: 1361 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.01 e-/ADU RON = 7.16 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.25/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay