*** MIDAS logfile (05SEPpl1.2) *** Mo, 16 Oct 2006 15:31:35 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 16 Oct 2006 15:32:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 061016 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.29 stddev = 5.51 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_061016 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1441.6 1442.3 1171.5 -0.08 -0.08 1.23 2 2.12 2912.5 2911.4 2327.1 0.23 0.20 1.25 3 4.25 5822.7 5821.9 4637.3 -0.10 -0.11 1.26 4 6.38 8746.9 8747.2 6955.3 -0.07 -0.06 1.26 5 9.57 13132.6 13133.2 10447.7 -0.03 -0.03 1.26 6 12.76 17532.2 17533.6 13969.9 0.05 0.06 1.25 7 15.95 21919.1 21918.7 17302.2 0.05 0.06 1.27 8 21.28 29232.4 29230.9 22862.0 0.05 0.05 1.28 9 26.59 36528.5 36527.4 29015.1 0.01 0.01 1.26 10 31.91 43783.0 43785.0 33319.6 -0.12 -0.11 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1868 to 56723 e- Average count rate measured: 1374 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 7.14 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.010 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.22/-0.11 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay