*** MIDAS logfile (05SEPpl1.2) *** Fr, 06 Oct 2006 14:43:39 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 06 Oct 2006 14:44:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 061006 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.02 stddev = 4.90 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_061006 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1529.4 1529.4 1246.0 -0.09 -0.08 1.23 2 2.12 3041.8 3041.5 2452.5 0.21 0.20 1.24 3 4.25 6037.5 6035.8 4805.6 -0.09 -0.11 1.26 4 6.38 9046.0 9046.2 7278.4 -0.05 -0.04 1.24 5 9.57 13555.5 13555.5 10703.9 -0.03 -0.03 1.27 6 12.76 18083.0 18082.8 14264.1 0.06 0.06 1.27 7 15.95 22596.0 22597.8 17889.5 0.06 0.07 1.26 8 21.27 30117.4 30117.2 27000.6 0.05 0.05 1.12 9 26.59 37620.0 37620.8 29271.2 0.00 0.00 1.29 10 31.91 45078.0 45076.7 34017.8 -0.13 -0.13 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1957 to 57668 e- Average count rate measured: 1413 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.05 e-/ADU RON = 6.27 +/- 0.23 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.023 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay