*** MIDAS logfile (05SEPpl1.2) *** We, 27 Sep 2006 18:16:29 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 27 Sep 2006 18:17:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.36 stddev = 4.55 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1498.1 1498.9 1186.7 -0.02 -0.03 1.26 2 2.13 2982.8 2982.8 2352.3 0.19 0.18 1.27 3 4.25 5917.9 5916.7 5091.5 -0.20 -0.21 1.16 4 6.38 8871.3 8870.5 6915.0 -0.14 -0.14 1.28 5 9.57 13297.9 13297.6 10305.9 -0.10 -0.10 1.29 6 12.76 17742.0 17744.2 13762.3 0.01 0.02 1.29 7 15.95 22180.0 22181.4 17167.7 0.05 0.06 1.29 8 21.27 29581.3 29582.0 22796.7 0.10 0.10 1.30 9 26.59 36968.9 36970.1 28097.9 0.10 0.10 1.32 10 31.91 44325.4 44324.0 33548.0 0.02 0.02 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1987 to 58764 e- Average count rate measured: 1388 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 6.04 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.020 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay