*** MIDAS logfile (05SEPpl1.2) *** We, 13 Sep 2006 16:30:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 13 Sep 2006 16:31:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060913 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 284.43 stddev = 5.26 FIERA temperature = 287.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060913 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1498.8 1499.1 1320.6 -0.02 -0.00 1.13 2 2.13 2970.1 2968.4 2352.1 0.15 0.13 1.26 3 4.25 5880.2 5879.5 4605.6 -0.21 -0.22 1.28 4 6.38 8809.3 8807.4 6845.1 -0.15 -0.16 1.29 5 9.57 13199.2 13199.5 10284.6 -0.11 -0.10 1.28 6 12.76 17610.0 17611.4 13576.6 0.02 0.02 1.30 7 15.95 22010.6 22012.0 18105.3 0.05 0.06 1.22 8 21.27 29353.1 29355.0 22899.2 0.10 0.12 1.28 9 26.59 36686.8 36685.5 27893.8 0.12 0.12 1.32 10 31.91 43979.6 43978.9 33405.3 0.04 0.04 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1969 to 57764 e- Average count rate measured: 1376 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.02 e-/ADU RON = 6.91 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.029 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay