*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 15:55:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 15:56:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060906 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.68 stddev = 5.25 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1418.1 1419.9 2211.6 -0.00 0.00 0.64 2 2.12 2866.8 2865.8 2254.5 0.14 0.09 1.27 3 4.25 5736.0 5736.2 4452.1 -0.22 -0.23 1.29 4 6.38 8623.4 8622.9 6734.7 -0.16 -0.16 1.28 5 9.57 12953.3 12953.4 10251.6 -0.10 -0.10 1.26 6 12.76 17300.4 17300.3 13487.6 0.02 0.03 1.28 7 15.95 21639.8 21639.6 16746.2 0.06 0.06 1.29 8 21.27 28875.7 28877.9 22230.0 0.10 0.12 1.30 9 26.59 36106.5 36105.5 27655.7 0.12 0.13 1.31 10 31.91 43300.6 43302.7 32662.9 0.05 0.06 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1898 to 57910 e- Average count rate measured: 1357 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 7.02 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.015 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay