*** MIDAS logfile (05SEPpl1.2) *** Th, 24 Aug 2006 20:42:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 24 Aug 2006 20:43:24 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060824 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 285.88 stddev = 5.95 FIERA temperature = 288.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060824 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1509.2 1508.2 1208.2 0.00 -0.01 1.25 2 2.12 3000.6 3000.2 2393.6 0.17 0.18 1.25 3 4.25 5952.8 5951.2 4674.7 -0.20 -0.22 1.27 4 6.38 8922.0 8921.8 7003.9 -0.15 -0.16 1.27 5 9.57 13375.5 13375.3 10409.3 -0.11 -0.10 1.28 6 12.76 17848.8 17848.1 13949.7 0.02 0.01 1.28 7 15.95 22313.2 22312.7 17179.8 0.06 0.06 1.30 8 21.27 29755.9 29757.7 22893.2 0.10 0.10 1.30 9 26.59 37188.9 37189.9 28921.4 0.10 0.10 1.29 10 31.91 44586.1 44588.0 34027.3 0.02 0.02 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1970 to 58212 e- Average count rate measured: 1396 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 7.77 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.021 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay