*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:45:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:46:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 286.19 stddev = 5.39 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1429.4 1430.1 1143.4 0.02 0.03 1.25 2 2.13 2884.7 2885.3 2267.0 0.16 0.18 1.27 3 4.25 5766.8 5766.9 4473.0 -0.22 -0.22 1.29 4 6.38 8666.6 8666.2 7101.0 -0.15 -0.16 1.22 5 9.57 13014.5 13014.3 10366.8 -0.12 -0.11 1.26 6 12.76 17383.4 17382.3 13775.2 0.02 0.01 1.26 7 15.96 21740.9 21738.7 16806.2 0.05 0.05 1.29 8 21.27 29009.2 29009.8 22294.2 0.10 0.10 1.30 9 26.59 36268.6 36268.4 27582.4 0.10 0.11 1.31 10 31.91 43496.1 43495.1 32812.6 0.04 0.03 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1898 to 57744 e- Average count rate measured: 1363 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.15 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.012 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.22 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay