*** MIDAS logfile (05SEPpl1.2) *** Th, 10 Aug 2006 14:28:52 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 10 Aug 2006 14:29:45 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060810 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 283.52 stddev = 5.23 FIERA temperature = 287.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060810 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1428.6 1429.2 1145.0 0.03 0.01 1.25 2 2.12 2885.4 2884.9 2325.1 0.17 0.13 1.24 3 4.25 5770.6 5769.7 4529.2 -0.22 -0.23 1.27 4 6.38 8673.8 8672.6 6774.6 -0.15 -0.16 1.28 5 9.57 13025.9 13026.5 10237.7 -0.12 -0.10 1.27 6 12.76 17399.9 17397.8 13646.6 0.02 0.02 1.28 7 15.95 21760.1 21760.5 16925.7 0.04 0.06 1.29 8 21.27 29036.6 29035.2 22301.3 0.10 0.11 1.30 9 26.59 36304.5 36304.8 27948.8 0.11 0.12 1.30 10 31.91 43536.9 43536.3 32718.6 0.03 0.05 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1890 to 57574 e- Average count rate measured: 1364 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 6.92 +/- 0.05 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = -0.013 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.23 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay