*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 18:42:58 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 18:43:50 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 283.96 stddev = 5.38 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1512.7 1512.5 1217.0 -0.03 0.04 1.24 2 2.12 2995.0 2994.5 2346.3 0.14 0.15 1.28 3 4.25 5930.8 5931.0 4675.8 -0.21 -0.21 1.27 4 6.38 8884.4 8883.3 7106.9 -0.15 -0.16 1.25 5 9.57 13313.5 13313.4 10393.7 -0.10 -0.11 1.28 6 12.76 17763.1 17762.5 13958.9 0.03 0.02 1.27 7 15.95 22201.0 22200.5 17193.1 0.06 0.05 1.29 8 21.27 29604.0 29604.3 22617.5 0.11 0.10 1.31 9 26.59 36995.3 36996.7 28270.7 0.11 0.10 1.31 10 31.91 44353.6 44352.5 33545.4 0.04 0.02 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2001 to 58681 e- Average count rate measured: 1388 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.12 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.029 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay