*** MIDAS logfile (05SEPpl1.2) *** Th, 20 Jul 2006 18:29:51 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 20 Jul 2006 18:30:59 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060720 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 282.58 stddev = 6.20 FIERA temperature = 288.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060720 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1467.2 1470.0 1209.6 0.09 0.03 1.21 2 2.13 2933.3 2935.2 2314.4 -0.10 -0.01 1.27 3 4.26 5851.8 5853.8 4602.7 -0.20 -0.14 1.27 4 6.38 8784.9 8783.1 6921.3 -0.06 -0.15 1.27 5 9.57 13179.2 13175.9 10301.4 -0.02 -0.09 1.28 6 12.77 17582.5 17596.0 13723.3 -0.01 0.08 1.28 7 15.96 21987.9 21985.4 17063.8 0.04 0.02 1.29 8 21.27 29340.5 29339.9 22417.0 0.14 0.12 1.31 9 26.60 36667.0 36669.6 28000.9 0.11 0.11 1.31 10 31.92 43965.0 43978.4 33413.0 0.02 0.05 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1937 to 58005 e- Average count rate measured: 1377 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 8.18 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.002 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.18 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay