*** MIDAS logfile (05SEPpl1.2) *** Su, 14 May 2006 16:41:34 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 14 May 2006 16:42:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060514 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.54 stddev = 4.73 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060514 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1574.7 1574.7 1285.4 -0.06 -0.07 1.23 2 2.13 3130.7 3130.0 2502.3 0.19 0.17 1.25 3 4.25 6215.1 6215.6 4954.1 -0.09 -0.08 1.25 4 6.38 9315.8 9314.8 7434.5 -0.03 -0.04 1.25 5 9.57 13959.0 13959.9 11085.8 -0.02 -0.02 1.26 6 12.76 18622.5 18622.4 14640.0 0.08 0.08 1.27 7 15.95 23269.6 23268.8 18270.4 0.07 0.07 1.27 8 21.27 31011.0 31011.0 24123.7 0.05 0.05 1.29 9 26.59 38729.1 38732.2 29742.0 -0.02 -0.01 1.30 10 31.91 46403.3 46405.8 35378.7 -0.16 -0.15 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2066 to 60869 e- Average count rate measured: 1455 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 6.21 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.022 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay