*** MIDAS logfile (05SEPpl1.2) *** Mo, 08 May 2006 22:50:21 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 08 May 2006 22:51:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060508 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.76 stddev = 6.05 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060508 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1576.6 1576.1 1285.3 -0.09 -0.11 1.23 2 2.13 3134.0 3133.9 2583.9 0.18 0.18 1.21 3 4.25 6221.4 6220.9 4959.8 -0.08 -0.09 1.25 4 6.38 9324.1 9323.1 7512.0 -0.03 -0.03 1.24 5 9.57 13970.8 13970.7 11066.1 -0.02 -0.02 1.26 6 12.76 18637.8 18636.9 15047.5 0.08 0.08 1.24 7 15.96 23286.2 23287.0 18813.2 0.07 0.08 1.24 8 21.28 31035.0 31033.9 24047.5 0.05 0.06 1.29 9 26.59 38760.8 38760.9 29813.6 -0.01 -0.00 1.30 10 31.91 46437.3 46439.2 34930.5 -0.16 -0.15 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2084 to 61400 e- Average count rate measured: 1456 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.02 e-/ADU RON = 8.01 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.022 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay