*** MIDAS logfile (05SEPpl1.2) *** Fr, 21 Apr 2006 17:41:48 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 21 Apr 2006 17:42:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060421 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.52 stddev = 5.61 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060421 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1574.4 1574.8 1276.9 -0.06 -0.05 1.23 2 2.13 3133.3 3132.5 2999.2 0.21 0.19 1.04 3 4.25 6222.0 6220.6 4982.6 -0.08 -0.09 1.25 4 6.38 9323.9 9324.2 7388.6 -0.04 -0.04 1.26 5 9.57 13972.9 13973.8 11035.1 -0.03 -0.03 1.27 6 12.76 18641.6 18642.1 14735.9 0.07 0.07 1.27 7 15.96 23294.7 23294.4 18331.5 0.07 0.07 1.27 8 21.28 31044.1 31046.7 24323.3 0.04 0.05 1.28 9 26.59 38774.2 38776.7 29936.9 -0.02 -0.01 1.30 10 31.91 46460.6 46459.9 35150.8 -0.16 -0.16 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2077 to 61293 e- Average count rate measured: 1457 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.40 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.020 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.20/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay