*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 19:24:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 19:25:18 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rD2 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 346.12 stddev = 5.79 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rD2_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1601.0 1600.3 1337.2 -0.10 -0.11 1.20 2 2.13 3182.6 3182.2 2566.4 0.20 0.18 1.24 3 4.25 6316.5 6316.2 5032.5 -0.09 -0.08 1.26 4 6.38 9464.6 9463.6 7535.1 -0.04 -0.04 1.26 5 9.57 14182.9 14182.3 11189.5 -0.02 -0.02 1.27 6 12.76 18917.9 18918.3 14804.4 0.07 0.08 1.28 7 15.96 23638.7 23638.3 18673.6 0.07 0.08 1.27 8 21.28 31503.2 31502.8 24520.9 0.06 0.06 1.28 9 26.59 39345.5 39347.1 30556.9 -0.01 0.00 1.29 10 31.91 47139.5 47139.4 36372.3 -0.15 -0.15 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2076 to 61140 e- Average count rate measured: 1478 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.00 e-/ADU RON = 7.51 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.023 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay